28
© Adam Equipment Company 2010
12.0 SERVICE
PARAMETERS
12.1 USING
“0000
TO
ENTER
THE
CALIBRATION
PARAMETER
Press the [Tare] key during the display counting when turned on. The
display will ask for a code number, “Pin on the Weight Display.
Enter the number 0000 when “Pin “ is displayed and then press
[Tare]. If another PIN number has been previously stored enter that
number instead.
The displays will show the first parameters, called “F1” “CAL”.
To select another parameter press the [U.Wt./Units] key to advance
through the parameters.
Press the [Tare] key to enter a parameter.
To exit a parameter, press the [Zero] key.
The Weight window will show the parameter number and the Unit
Price and Count windows will show the word describing the function.
When a parameter is entered by pressing the [Tare] key, the displays
will guide you through the parameter selected and the options
available.
The parameters available are:
“F1” CAL”
To enter the Calibration
“F2” “Pin To set a new Pin number for access
“F3” “Unit” Enable Weighing Units
Vue de la page 29